HDI Currents
The current draws using benchtop supplies are listed below. These are with CLOCK A enabled and CLOCK B disabled. The voltages were set using the senselines (AVDD=5 V, AVDD2=2.2V, DVDD=5V). The measurement accuracy is of order 10 mA.
HDI |
Side |
DVDD |
AVDD |
AVDD2 |
---|---|---|---|---|
H1-13 |
n |
420 mA |
320 mA |
120 mA |
H1-13 |
p |
440 mA |
320 mA |
120 mA |
H1-29 |
n |
480 mA |
350 mA |
110 mA |
H1-29 |
p |
470 mA |
330 mA |
110 mA |
The currents after bonding to the silicon for HDI H1-29 (DFA D02F-4) using the benchtop supplies, were consistent with the values before bonding.
HDI H1-13, (DFA D02B-8) after bonding to the silicon, was used during some of the tests of the new CAEN power supply module. The CAEN supply measures the currents with an accuracy of order 1 mA. The tables below give the measured currents for when the chip is idle and during a calibration run. Note that the DVDD current drops about 70 mA during a calibration run, while the analog currents remain the same. A similar drop was observed with HDI H1-29 (DFA D02F-4).
HDI / DFA |
Side |
DVDD |
AVDD |
AVDD2 |
---|---|---|---|---|
H1-13 / D02B-8 |
n |
461 mA |
343 mA |
105 mA |
H1-13 / D02B-8 |
p |
470 mA |
330 mA |
124 mA |
HDI / DFA |
Side |
DVDD |
AVDD |
AVDD2 |
---|---|---|---|---|
H1-13 / D02B-8 |
n |
392 mA |
342 mA |
105 mA |
H1-13 / D02B-8 |
p |
398 mA |
330 mA |
123 mA |
Silicon Currents
The table below gives the operating voltage, p-side bias current, and p-side edge guard current for the two detectors. Typically, a Layer 2 DFA will draw from 1 to 2 uA on both the bias and the edge guard. However, the currents for DFA D02F-4, which is connected to HDI H1-29, are unstable. The detector initially drew 6.3 uA and 5.6 uA for the p-side bias and edge guard respectively. After about a half hour at 40 V, the currents were above 50 uA.Detector | Voltage | Bias-p Current | Guard-p Current |
D02B-8 (H1-13) | 40 V |
1.3 uA |
1.1 uA |
D02F-4 (H1-29) | 40 V |
From 6 uA to 80 uA |
From 6 uA to 80 uA |